- New
Examination Light
DR MACH LED 130F
DR MACH LED 130F
DR MACH LED 130F
This light will be used illumination during delivery, minor surgical procedures and general examination of patients.
The examination light uses a single bulb.
The field size is no less than 50 cm at 1m distance.
The focal length is no less than 70 cm.
The illumination level is no less than 50,000 lux at 1 meter.
The examination light rotates 360° from a fixed point.
The examination light is on a mobile stand with goose neck arm
The stand is height adjustable and mounted on a supporting base
The arm is less than 50cm and the overall height not less than 170cm
Casters
Stability
The supply includes 2 sets of spare bulbs
Environmental factors
The unit is capable of being stored continuously in ambient temperature of 180C to 50 0C and relative humidity of 15-95%
Meets IEC-60601-1-2: 2001 or equivalent for General Requirements of Safety for Electromagnetic Compatibility.
The unit is capable of operating continuously in ambient temperature of 180C to 40 0C and relative humidity of 15-95%
Power supply
The unit operates satisfactorily on power input of 220-240VAC, 50Hz with tolerance of ±20%
Where necessary, a suitable Servo controlled Stabilizer should be provided to protect the unit against power fluctuation
The unit is fitted with a resettable over current breaker for protection for protection against sudden power surge
The unit is fitted with appropriate cable length not less than 3m and fitted with a British Standard plug top
Where necessary, a suitable UPS with maintenance free batteries for minimum one three back up should be supplied with the system.
Standards and Safety
Certified to be compliant with appropriate IEC60601.2.12-01 Medical Electrical Equipment as well as the particular requirements for the unit
FDA or CE approved product. Where FDA approval applies, the power supply should be compatible with input voltage of 220 to 240V AC without external step down transformer required
Certified to be compliant with applicable ISO standards for the unit
Meets IEC 529 Level 3 (IP3X) (spraying water) for enclosure protection, water ingress.
Has the necessary equipments recommended by the manufacturer to carry out preventive maintenance test as per guidelines provided in the service/maintenance manual.
The unit is provided with a comprehensive warranty for 2 years minimum.
Documentation
The equipment is supplied with a certificate of calibration and inspection from factory as well as independent certification body
List of tools and equipments required for providing calibration and routine maintenance as per manufacturer documentation in service / technical manual.
User Manual in English
Service/maintenance manual with detailed information for servicing, troubleshooting and repairs
A logbook with checklist for daily, weekly, monthly and quarterly maintenance should be provided
List of important spare parts and accessories with their part numbers and costing must be provided.